Standards:
In accordance with or exceed the requirements of the following standards: ISO 7500-1, ASTM E4, EN10002-2, BS 1610, DIN 51221, ISO6892.
1. Complete computer controlled: The whole measuring and controlling system adopts specific PC control card
Used for testing machines, realizing the data zero and plus adjustment, which has very high reliability.
2. Supporting multi-transducers
Realizing the database management of the test data which are stored according to the standard format;
Facilitating other software to analyze and transfer.
3. Perfect programming by auto program control, every control mode can smoothly shift to another one.
Fulfilling the test requirement of all kinds of materials with every test standard home and abroad.
4. Control software has the auto-adjusting function of test hardness, which assures that the system works with
Every kind of specimen hardness.
5. Perfect graphic function realizing the arbitrarily magnifying, decreasing, equaling, adding, indicative display and
Print of all kinds of test curves, the test point searching as well as the simultaneous display and print of several
Kinds of test curves. Data processing supposes self-disposing and input disposing of graphic human computer Interaction, which facilitates the check and contrast of the test result.
6. The user can self -define the output of the test report, which makes the report format have very high flexibility.
7. Modularization design facilitates the software upgrading, function spreading as well as the second development.
8. All the crucial parts were processed by our owen processing center which was imported from Japan to guarantee The high accuracy of machine.
In accordance with or exceed the requirements of the following standards: ISO 7500-1, ASTM E4, EN10002-2, BS 1610, DIN 51221, ISO6892.
1. Complete computer controlled: The whole measuring and controlling system adopts specific PC control card
Used for testing machines, realizing the data zero and plus adjustment, which has very high reliability.
2. Supporting multi-transducers
Realizing the database management of the test data which are stored according to the standard format;
Facilitating other software to analyze and transfer.
3. Perfect programming by auto program control, every control mode can smoothly shift to another one.
Fulfilling the test requirement of all kinds of materials with every test standard home and abroad.
4. Control software has the auto-adjusting function of test hardness, which assures that the system works with
Every kind of specimen hardness.
5. Perfect graphic function realizing the arbitrarily magnifying, decreasing, equaling, adding, indicative display and
Print of all kinds of test curves, the test point searching as well as the simultaneous display and print of several
Kinds of test curves. Data processing supposes self-disposing and input disposing of graphic human computer Interaction, which facilitates the check and contrast of the test result.
6. The user can self -define the output of the test report, which makes the report format have very high flexibility.
7. Modularization design facilitates the software upgrading, function spreading as well as the second development.
8. All the crucial parts were processed by our owen processing center which was imported from Japan to guarantee The high accuracy of machine.
Specifications | WDW-600E/500E | WDW-300E | WDW-200E | WDW-100E | WDW-50E | WDW-30E/20E/10E | WDW-5E/2E/1E | |
Capacity(kN) | 600/500 | 300 | 200 | 100 | 50 | 30/20/10 | 5/2/1 | |
Load frame Stiffness (kN/mm) | 1000 | 600 | 400 | 300 | 250 | 80 | 30 | |
Load range | 0.4%-100% of the max load | 2%-100% | ||||||
Accuracy of test load | ≤±0.5% | |||||||
Test Space | Single / Double test space | Single | ||||||
Precision of Ball Screws | 16μm/300mm E level | |||||||
Accuracy of deformation | <±1% within the 2%-100% full range of the extensometer (accuracy level±0.5%) | |||||||
Crosshead stroke accuracy | 0.001mm | |||||||
Accuracy of indication value of test load | <±1% (accuracy level±0.5%) | |||||||
Resolution of load | 1/200000 of the max load force | |||||||
Scope of deformation measure (normal extensometer) |
2%-100% FN | |||||||
Accuracy Indication of deformation (normal extensometer) | within ±1% of indication value | |||||||
Scope of deformation measure (High deformation extensometer) |
10mm-800mm | |||||||
Accuracy Indication of deformation (High deformation extensometer) |
within the ±1% of the value | |||||||
Resolution of crosshead stroke | 0.001mm | |||||||
Adjustment scope of test speed under Load control mode | 0.005-5%FN/S | |||||||
Accuracy of test speed under Load control mode | Test Speed<0.05%FN/s,within the ±2% of the preset value; Test Speed≥0.05%FN/s, within the ±0.5% of the preset value | |||||||
Adjustment Scope of deformation rate | 0.005-5%FN/S | |||||||
Accuracy of deformation rate | Test Speed<0.05%FN/s,within the ±2% of the preset value, while Test Speed≥0.05%FN/s,within the ±0.5% of the preset value | |||||||
Adjustment scope of stroke speed | 0.005mm/min-250mm/min | 0.005mm/min-500mm/min | ||||||
Accuracy of stroke speed | Test speed<0.01mm/min, within the ±1.0% of preset value, while test speed≥0.01mm/min, within the ±2% of the preset value | |||||||
Scope of the consistent load deformation and displacement control | 0.5%-100%FN/s | |||||||
Accuracy of the consistent load deformation and displacement control |
preset value≥10%FN, within the ±0.1% of preset value; preset value<10%FN, within the ±1% of preset value |
|||||||
Length of the test space(mm) | 600 | 600 | 600 | 600 | 600 | 800 | 700 | |
Width of the test space(mm) | 650 | 600 | 600 | 600 | 575 | 370 | 400 | |
Dimension(mm) | 1150x770x2817 | 1100x770x2817 | 1100x770x2685 | 1010x750x2225 | 945x654x2266 | 686x525x1880 | 610x480x1285 | |
Weight(kg) | 2800 | 1560 | 1560 | 1100 | 700 | 250 | 100 | |
Power Supply | AC380v±10%, 50/60Hz, Three-phase Five-wire | AC220v±10%, 50/60Hz | ||||||
Power | 5 | 3 | 1.5 | 0.75 | 0.4 | |||
Type of machine | Floor type | Table type |